Key Specifications
-
Measurement Range
ORP: –1,500 to 1,500 mV
Application B
• pH: 0 to 14
Application H
• pH: 0 to 12 -
Operating Temperature
Applications B & H:
0 to 140°C (32 to 284°F)
TB Version:
0 to 140°C (32 to 284°F)
TU/TP Version (pressurized reference):
0 to 140°C (32 to 284°F) (140°C/284°F for sterilization only)
Max. 100°C (212°F) continuous operation due to pressure drop at T>100°C (212°F) -
Operating Pressure
Application B: 0.8 to 14 bar (11.6 to 203 psi) absolute
Application H: 0.8 to 7 bar (11.6 to 101.5 psi) absolute
Applications
The Memosens CPS76E is a rugged solution for concurrent pH and ORP measurement. Its contamination-resistant reference system delivers stable performance in challenging environments with pollutants, toxic substances, or low conductivity. Featuring Memosens 2.0 digital technology, it combines process reliability with user-friendly operation. The sensor is moisture-resistant, supports lab calibration, and stores extensive calibration/process data for predictive maintenance.
Ideal for chemical processes and monitoring applications with:
-
Rapid pH/ORP fluctuations
-
High concentrations of electrode poisons like H2S
Certified for hazardous areas (ATEX, IECEx, CSA C/US, NEPSI, Japan, INMETRO) in zones 0, 1, and 2.
Advantages
-
Memosens 2.0 stores comprehensive calibration and process data for trend analysis, supporting predictive maintenance and IIoT integration.
-
Simultaneous pH, ORP, and rH measurement (in rH mode) enhances process visibility and control.
-
Optional pressurized reference maintains accuracy in challenging media like dispersions.
-
Patented ion trap protects the reference system from poisoning, extending sensor life.
-
Versatile installation options including inverted mounting.
-
Inductive signal transmission ensures maximum process safety.
-
Reduces operational costs through extended sensor life and minimized downtime.