Key Specifications
-
Maximum measurement accuracy
Liquid mass flow: ±0.10 %
Liquid volume flow: ±0.10 %
Gas mass flow (Tantalum only): ±0.50 %
Liquid density: ±0.0005 g/cm³ -
Measurement range
0 to 70,000 kg/h (0 to 2,570 lb/min)
-
Operating temperature range
Tantalum: –50 to +150 °C (–58 to +302 °F)
Zirconium: –50 to +205 °C (–58 to +401 °F) -
Maximum working pressure
PN 40, Class 300, 20K
-
Contact materials
Measurement tube: Tantalum 2.5W; 702 (UNS R60702)
Connections: Tantalum; 702 (UNS R60702)
Applications
The precision-engineered Promass H is designed for applications demanding superior corrosion resistance, ensuring reliable performance with chemically aggressive media. Featuring the advanced remote transmitter Promass H 500, it offers exceptional installation versatility and operational security in challenging conditions. Heartbeat Technology provides continuous process monitoring for enhanced safety.
-
Measurement principle unaffected by fluid characteristics including viscosity or density
-
Precision measurement of liquids and gases in highly corrosive environments
Product Features
-
Tantalum or Zirconium measurement tube
-
Size range: DN 8 to 50 (3/8 to 2″)
-
Maximum media temperature: +205 °C (+401 °F)
-
Remote configuration with up to 4 I/O channels
-
Touch-enabled backlit display with WLAN connectivity
-
Included standard sensor-transmitter cable
Advantages
-
Exceptional protection for corrosive media – corrosion-resistant wetted components
-
Reduced instrumentation needs – simultaneous measurement of flow, density and temperature
-
Compact installation – no straight pipe requirements
-
Comprehensive data access – flexible I/O combinations and Ethernet communication
-
Simplified configuration – customizable I/O functionality
-
Built-in verification – Heartbeat Technology