Key Specifications
-
Measurement Range
Trace: 0–5 mg/l HOCl
Standard: 0–20 mg/l HOCl
High: 0–200 mg/l HOCl -
Operating Temperature
0–55 °C (32–130 °F), non-freezing conditions
-
Operating Pressure
Maximum 1 bar (14.5 psi)
-
Measurement Principle
Sealed membrane-covered sensor
Free chlorine reduction at the cathode
Applications
The Memosens CCS51D is a durable, low-maintenance free chlorine sensor. It delivers precise measurements with excellent stability in drinking water, swimming pools, and industrial water systems, ensuring reliable disinfection monitoring even at low flow rates. With its rapid response capability, it enables efficient process management. Note: The CCS51D is not the newest Memosens model. For details on the advanced Memosens CCS51E sensor.
The Memosens CCS51D is designed for skid manufacturers and end-users, providing dependable measurements in:
-
Drinking water – ensuring effective disinfection
-
Process water – maintaining hygiene in packaging and bottling
-
Utilities – detecting free chlorine presence
-
Pool water – optimizing disinfectant dosing
Advantages
-
Versatile sensor options for all applications, from trace levels up to 200 mg/l free chlorine.
-
Rapid response (t90 < 25 s) ensures real-time monitoring and quick adjustments for optimal process control.
-
Enhanced process reliability with stable, long-term measurements for consistent monitoring and precise disinfectant dosing.
-
Low-maintenance amperometric design reduces operational costs compared to colorimetric systems.
-
Minimized downtime with quick sensor replacement: pre-calibrate in the lab and install with plug & play simplicity.
-
Seamless integration with Liquiline transmitters for combined analysis of key parameters like pH and ORP.