Measuring principle
Conductive
-
Application
Calibration of conductivity measuring loops in ultra-pure water applications in the pharmaindustry; calibration of quality-relevant process measurements in the ultra-pure water range up to 20µS/cm
-
Characteristic
factory calibrated measuring case with certificate for comparison measurement in ultra-pure water up to 20µS/cm
-
Measuring range
0,04µS/cm – 20µS/cm
-
Measuring principle
conductiv conductivity measurement
-
Design
compact case includes factory calibrated conductivity loop and flow-through assembly
-
Material
case : injection molded high performance resin
Conductivity sensor CLS15D: SS 316 L, PES-GF20
flow-through assembly: PVDF -
Dimension
20.9″ x 17.4″ x 8.5″
-
Process temperature
0.. 100C
32 .. 210F -
Process pressure
87 psi
-
Ingress protection
IP67 at closed case without power cord
-
Additional certifications
factory calibration certificate according ASTM 5391-93, traceable to SRM from NIST and DaakS
Field of application
Conducal CLY421 is the ideal calibration tool for conductivity measuring devices in ultrapure water applications. It offers certified comparison measurement and full traceability of your instruments’ verification and calibration. Furthermore, the tool is fully compliant with the requirements of the pharmaceutical industry. Conducal comes in a robust, portable hard-top case with battery-powered Liquiline transmitter, allowing flexible calibration in the field.
Conducal CLY421 is a portable calibration and verification tool for quality-relevant conductivity measurements. It is suitable for all ultrapure water applications up to 20 µS/cm:
-
Life science industry
-
Food industry
-
Semiconductor industry
Its traceability to national standards (NIST and DAkkS) fulfills the highest regulatory requirements.
Benefits
-
Factory calibration traceable to standard reference material by NIST and DAkkS
-
Factory calibration of Conducal according to ASTM D-5391
-
Flow assembly with adjusting and monitoring functions according to ASTM D-5391
-
Accurate calibration by direct use of process sample and comparison with the reference system
-
Rapid setup of comparison measurement with practical flow assembly
-
Flexible calibration in the field with carrying case and battery-powered Liquiline transmitter